For measurements up to 30mm (1.18 inches), the TLAser230 micrometer provides diameter, ovality, and position data with resolution of .000001 inch and repeatability of .00001 inch. With exceptional accuracy, compact design, and robust service, it's the best choice for your most demanding applications.
The TLAser230 micrometer can continuously measure product diameters and product position in two planes at right angles to each other. A maximum gauging rate of 8,000 measurements per second enables the TLAser230 to check diameter and ovality while also detecting lumps and neck-downs, to assure the most thorough quality inspection possible.
Incorporating a mirror-based design, the rugged TLAser230 Micrometer features a heavy aluminum body with sealed metal covers for lasting performance in harsh industrial environments. An optional air purge mounts easily to the gauge.
The TLAser230, like all LaserLinc micrometers, is designed to serve only one purposemeasure the product's dimensions and route them to a PC. The TLAser230 micrometer operates with the popular TLAser400 interface card and related software to control product quality more economically and effectively by capitalizing on the enormous processing power and flexibility of today's PCs.
|
For LaserLinc’s other dual-axis micrometers, please visit these pages.
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Brochures and Mechanical Drawings
LaserLinc dual-axis micrometer brochure (including TLAser230, TLAser260, and TLAser2120).
|
|
|
|
|
Measurement
Range
|
0.007˝ 1˝
(0.17mm 25.4mm)
|
|
|
|
Maximum
Measurement Size
|
1.18˝
(30mm)
|
|
|
|
Measurements
per second
|
3,200 (1,600 per axis) standard /
8,000 (4,000 per axis) optional
|
|
|
|
Resolution
|
.000001˝ (.025µm)
|
|
|
|
Repeatability*
|
|
|
Single Scan
|
±0.002˝ (±0.0508mm)
|
|
|
|
Two Second
|
±0.00001˝ (±.25µm)
|
|
|
|
Positional
Error
|
±0.0001˝ (±2.54µm)
within center 1" (25.4mm)
|
|
|
|
Approx.
Dimensions
(H x W x D)
|
10.75˝ x 13.25˝ x 2.15˝
(273mm x 337mm x 55mm)
|
|
|
|
Approx.
Weight
|
20 lbs
(9kg)
|
|
|
|
Power
requirements
from PC
|
+12VDC 200mA, +5VDC 150mA,
-5VDC 150mA
|
|
|
Part aperture is 3˝ x 4˝ (without air purge).
Run up to four TLAser230 micrometers per PC.
Laser Class II visible red laser diode wavelength is 675nm.
|
|
|
|